Test Code 62871039 D Dimer Quant
Performing Laboratory
Torrance Memorial Laboratories
Specimen Requirements
Specimen Type: Whole blood
Container/Tube: Blue top (citrate)
Specimen Volume: Tube must be filled
Collection Instructions:
1. Tube must be filled to the fill line.
2. If using butterfly, discard tube must be drawn first.
3. Spin immediately.
TML processing instructions: Platelet-poor plasma preparation
Specimen Transport Temperature
Ambient
Storage Temp Stability:
Ambient: 4 hours
Refrigerated: 4 hours
Frozen: 2 weeks
Methodology
Photometric
Reference Range
≤500 FEU ng/mL
The D-Dimer Hemosil HS 500 assay is for the quantitative determination of D-Dimer in plasma using an enhanced immunotubidimetric method. The results are reported in fibrinogen-equivalent units (FEU). D-Dimer levels can be increased in the following conditions: disseminated intravascular coagulation, cancers, hemorrhages, post-operative period, and infections.
Using a cut-off of 500 FEU ng/mL the assay can be used to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) with 100% sensitivity in conjunction with a clinical pretest probability assessment.
Reference ranges provided on this web site are for guidance only, and may not reflect the most recent changes. Refer to laboratory reports for current data.
Day(s) Test Set Up/TAT
Performed:
Daily
Available STAT
TAT:
STAT: < 1 hr
Routine: < 4 hrs
Aliases
None
CPT Code
85378