Sign in →

Browse by Name

  1. A
  2. B
  3. C
  4. D
  5. E
  6. F
  7. G
  8. H
  9. I
  10. J
  11. K
  12. L
  13. M
  14. N
  15. O
  16. P
  17. Q
  18. R
  19. S
  20. T
  21. U
  22. V
  23. W
  24. X
  25. Y
  26. Z
  27. #
 

Mayo Clinic Laboratories

Test Code 221738482 Circulating Anticoagulant Screen

Important Note

TML Outreach Instruction:  Perform expedite processing

Performing Laboratory

Torrance Memorial Laboratories

Specimen Requirements

Submit the following specimen:   

 

Specimen Type:  Whole blood

Container/Tube: Minimum (2) 2.7 mL blue top tubes 

Specimen Volume: Tube must be filled to its intended whole blood capacity; minimum requirement 3 mL of plasma

Collection Instructions:

1.  Tube must be filled to fill line

2.  Use a 21 gauge needle

 

Preferred specimen:  < 4 hours, OK to freeze platelet poor plasma

Unacceptable specimens:  Non blue top, partially filled blue top, hemolyzed specimen, clotted sample, > 4 hours old

Specimen Transport Temperature

Ambient

 

Storage Temperature Stability:

Room Temperature:  4 hours

Frozen:  14 days (platelet poor plasma)

If specimen cannot be delivered within 4 hours, the specimen must be double spun and plasma frozen.

Methodology

Werfen ACLtop 550

Reference Range

N/A

Day(s) Test Set Up/TAT

Set Up:

Daily (on demand)

Aliases

Mixing Study for Coagulation

CPT Code

85611